Mechanical test fixtures designed to ISO and FDA-referenced standards. Each fixture ships with a validated test protocol and reference data. You're not just buying a rig, you're buying a test method.
4×4 base plate with a fixed pin gage for kink resistance testing per FDA guidance. One pin diameter per test; the collet set holds pin gages of different diameters, so you change the bend diameter by swapping the pin rather than the fixture.
Standard: FDA-recommended kink test method
6×6 base plate with a stepped mandrel, giving a graduated series of bend diameters in a single part. Find the diameter at which a device kinks without swapping pins between runs.
Standard: FDA-recommended kink test method
Flat-plate kink fixture for compliance with BS EN 13868. Universal screw mount fits most universal testing machines on the market.
Standard: BS EN 13868
Block of U-shaped tracks at graduated radii, with a clear top plate for visual inspection. The groove supports the catheter wall the way a vessel or guide sheath would, and lets you test short or stiff distal segments.
Standard: FDA-recommended